2470 High Voltage SourceMeter Instrument Reference Manual
Section 14: TSP command reference
2470-901-01 Rev. A /
May
2019
14-253
Reset the instrument.
Reset trigger timer 4.
Set trigger timer 4 to have a 0.5 s delay.
Set the stimulus for trigger timer 4 to be the notify 8 event.
Set the trigger timer 4 stimulus to off.
Set the timer event to occur when the timer delay elapses.
Set the trigger timer 4 count to 20.
Enable trigger timer 4.
Clear the trigger model.
Set trigger model block 1 to clear the buffer.
Set trigger model block 2 to generate the notify 8 event.
Set trigger model block 3 to wait for the trigger timer 4 to occur.
Set trigger model block 4 to make a measurement and store it in default buffer 1.
Set trigger model block 5 to repeat the trigger model 20 times, starting at block 3.
Start the trigger model.
Wait until all commands are complete.
Print the number of entries in default buffer 1.
Output:
20
Also see
(on page 14-251)
(on page 14-256)
trigger.timer[N].delay
This attribute sets and reads the timer delay.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger timer
N
reset
Configuration script
10e-6 (10 µs)
Usage
interval
= trigger.timer[
N
].delay
trigger.timer[
N
].delay =
interval
interval
Delay interval in seconds (8 µs to 100 ks)
N
Trigger timer number (1 to 4)