2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-115
:SYSTem:POSetup
This command selects the defaults that are used when you power on the instrument.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
RST
Usage
:SYSTem:POSetup <name>
:SYSTem:POSetup?
<name>
Which setup to restore when you power on the instrument:
Power on to
*RST
defaults:
RST
Stored setup 0:
SAV0
Stored setup 1:
SAV1
Stored setup 2:
SAV2
Stored setup 3:
SAV3
Stored setup 4:
SAV4
Details
When you select
RST
, the instrument restores settings to their default values when the instrument is
powered on.
When you select a
SAV
option, the settings in the selected saved setup are applied when the
instrument is powered on. The settings are saved using the
*SAV
command.
Example
SYST:POS SAV1
Set the instrument to restore the settings that are saved in the
stored setup 1 when the instrument is powered on.
Also see
(on page 12-9)