Section 2: Installation
2470 High Voltage SourceMeter Instrument Reference Manual
2-40
2470-901-01 Rev. A /
May
2019
To verify which command set is selected from a remote interface:
Send the command:
*LANG?
To change to the SCPI command set from a remote interface:
Send the command:
*LANG SCPI
Reboot the instrument.
To change to the TSP command set from a remote interface:
Send the command:
*LANG TSP
Reboot the instrument.
System information
You can get the serial number, firmware build, detected line frequency, calibration verify date,
calibration adjust date, and calibration adjust count information from the instrument.
To view the version and serial number information from the front panel:
1.
Press the
MENU
key.
2.
Under System, select
Info/Manage
.
The firmware version and serial number are displayed at the top of the screen.
To view the calibration information from the front panel:
1.
Press the
MENU
key.
2.
Under System, select
Calibration
.
The adjust date, adjust count, and calibration date are displayed.
To view the line frequency information from the front panel:
1.
Press the
MENU
key.
2.
Under System, select
Settings
.
3.
Scroll down to display the line frequency.