Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-60
2470-901-01 Rev. A /
May
2019
If the NPLC setting is less than 0.2 PLC, sending autozero once can result in delay of more than a
second.
Example
FUNC "VOLT"
AZER:ONCE
Do a one-time refresh of the reference and zero measurements
for the voltage function.
Also see
Automatic reference measurements
(on page 4-42)
[:SENSe[1]]:<function>:AZERo[:STATe]
(on page 12-46)
[:SENSe[1]]:CONFiguration:LIST:CATalog?
This command returns the name of one measure configuration list that is stored on the instrument.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:CATalog?
Details
You can use this command to retrieve the names of measure configuration lists that are stored in the
instrument.
This command returns one name each time you send it. This command returns an empty string when
there are no more names to return. If the command returns an empty string the first time you send it,
no measure configuration lists have been created for the instrument.
Example
CONF:LIST:CAT?
Send this command to retrieve the name of one measure
configuration list. To get all stored lists, send it again until it
returns an empty string.
Also see
[:SENSe[1]]:CONFiguration:LIST:CREate
(on page 12-61)