2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-67
Also see
(on page 12-4)
(on page 12-120)
(on page 12-188)
[:SENSe[1]]:FUNCtion[:ON]
This command selects the active measure function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
CURR
Usage
[:SENSe[1]]:FUNCtion[:ON] "<function>"
[:SENSe[1]]:FUNCtion[:ON]?
<function>
A string that contains the measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
Details
Set this command to the type of measurement you want to make.
Reading this command returns the measure function that is presently active.
Example
:FUNC "VOLTage"
Make the voltage measurement function the active function.
Also see
Making resistance measurements
(on page 4-25)