Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-284
2470-901-01 Rev. A /
May
2019
For the
termSequence
parameter, use the same values listed in the table above for type. There are
four possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns
(CR or 0x13). For TSP-enabled devices, the default is
tspnet.TERM_LF
. For devices that are not
TSP-enabled, the default is
tspnet.TERM_CRLF
.
Example
deviceID = tspnet.connect("192.0.2.1")
if deviceID then
tspnet.termination(deviceID, tspnet.TERM_LF)
end
Sets termination type for IP address
192.0.2.1 to
TERM_LF
.
Also see
(on page 14-277)
(on page 14-278)
tspnet.timeout
This attribute sets the timeout value for the
tspnet.connect()
,
tspnet.execute()
, and
tspnet.read()
commands.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
No
Restore configuration
Instrument reset
Power cycle
Configuration script
20.0 (20 s)
Usage
value
= tspnet.timeout
tspnet.timeout =
value
value
The timeout duration in seconds (1 ms to 30.0 s)
Details
This attribute sets the amount of time the
tspnet.connect()
,
tspnet.execute()
, and
tspnet.read()
commands will wait for a response.
The time is specified in seconds. The timeout may be specified to millisecond resolution, but is only
accurate to the nearest 10 ms.
Example
tspnet.timeout = 2.0
Sets the timeout duration to 2 s.
Also see
(on page 14-277)
(on page 14-279)
(on page 14-281)