Section 4: Sourcing and measuring
2470 High Voltage SourceMeter Instrument Reference Manual
4-92
2470-901-01 Rev. A /
May
2019
5.
Select
Index Details
to see some of the settings stored in configuration index 1.Refer to
Instrument settings stored in a measure configuration list
(on page 4-86) for information about all
settings that are stored.
6.
Select
HOME
to return to the home screen.
Figure 81: Configuration list details
Storing configuration index 2
(on page 4-91), and continue as follows to
configure the instrument for configuration index 2.
Using the front panel to configure the instrument:
Change the following instrument settings to configure the instrument with the settings you want to
save for configuration index 2:
•
Set the measure range to 100 mA
•
Set NPLC to 2
Using the front panel to store the active settings to configuration index 2:
1.
Return to the configuration list menu.
2.
Select
MyMeasList
.
3.
Select
System to Index
. This saves the active system settings to the configuration index.
4.
Select
Index Details
to see some of the settings stored in configuration index 2. See
settings stored in a measure configuration list
(on page 4-86) for information about the settings
that are stored.