2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-117
:SYSTem:VERSion?
Query the present SCPI version.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:VERSion?
Details
This query command returns the SCPI version.
Example
SYSTem:VERSion?
Query the version. An example of a return is:
1996.0
Also see
None
TRACe subsystem
The TRACe subsystem contains commands that control the reading buffers.
:TRACe:ACTual?
This command contains the number of readings in the specified reading buffer.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:TRACe:ACTual?
:TRACe:ACTual? "<bufferName>"
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
Details
This command returns the number of readings stored in the buffer.
Example
TRACe:MAKE "testData", 200
COUN 10
MEASure:CURRent? "testData"
Creates 200 element reading buffer named
testData
.
Set the measurement count to
10
.
Set the measurement function to current. Make readings, and
store the readings in
testData
. Returns the tenth
measurement reading after taking all ten readings.