Section 12: SCPI command reference
2470 High Voltage SourceMeter Instrument Reference Manual
12-28
2470-901-01 Rev. A /
May
2019
The binary representation of the value indicates the output pattern to be written to the I/O port. For
example, a value of 63 has a binary equivalent of 111111 (all lines are set high); a
data
value of 42
has a binary equivalent of 101010 (lines 2, 4, and 6 are set high, and the other three lines are
set low).
An instrument reset does not affect the present states of the digital I/O lines.
All six lines must be configured as digital control lines. If not, this command generates an error.
Example
:DIG:WRIT 63
Sets digital I/O lines 1 through 6 high (binary 111111).
Also see
(on page 8-24)
Digital I/O port configuration
DISPlay subsystem
This subsystem contains commands that control the front-panel display.
:DISPlay:CLEar
This command clears the text from the front-panel USER swipe screen.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:DISPlay:CLEar
Example
DISP:CLE
DISP:SCR SWIPE_USER
DISP:USER1:TEXT "Batch A122"
DISP:USER2:TEXT "Test running"
Clear the USER swipe screen.
Display the USER swipe screen.
Set the first line to read "Batch A122" and the second line to
display "Test running".