Technical data
Proline Promag H 500 PROFIBUS DP
220
Hauser
0
0
0
0
-20
-40
-40
20
20
40
100
60
140
[°F] [°C]
40 60 80 100
100
120 140 160
200
300
180
360
-20
-40
-40
T
A
T
F
[°F]
[°C]
A0027450
41
Promag 500
T
A
Ambient temperature range
T
F
Fluid temperature
Conductivity
≥ 5 μS/cm for liquids in general. Stronger filter damping is required for very low
conductivity values.
Proline 500
The necessary minimum conductivity also depends on the cable length .
Pressure-temperature
ratings
An overview of the pressure-temperature ratings for the process connections is
provided in the "Technical Information" document
Pressure tightness
Liner: PFA
Nominal diameter
Limit values for absolute pressure in [mbar] ([psi]) for medium temperatures:
[mm]
[in]
+25 °C
(+77 °F)
+80 °C
(+176 °F)
+100 °C
(+212 °F)
+130 °C
(+266 °F)
+150 °C
(+302 °F)
2 to 150
¹⁄₁₂ to 6
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
Flow limit
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor.
The optimum velocity of flow is between 2 to 3 m/s (6.56 to 9.84 ft/s). Also match the
velocity of flow (v) to the physical properties of the fluid:
• v < 2 m/s (6.56 ft/s): for low conductivity values
• v > 2 m/s (6.56 ft/s): for fluids producing buildup (e.g. milk with a high fat content)
A necessary increase in the flow velocity can be achieved by reducing the sensor
nominal diameter.
For an overview of the full scale values for the measuring range, see the "Measuring
range" section → 207
Pressure loss
• No pressure loss occurs as of nominal diameter DN 8 (5/16") if the sensor is installed in
a pipe with the same nominal diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545