Proline Promag H 500 PROFIBUS DP
Electrical connection
Hauser
39
1
2
3
4
5
6
7
a
b
A0029151
15
Cable cross-section
a
Electrode cable
b
Coil current cable
1
Core
2
Core insulation
3
Core shield
4
Core jacket
5
Core reinforcement
6
Cable shield
7
Outer jacket
Operation in zones of severe electrical interference
The measuring system meets the general safety requirements → 230 and EMC
specifications → 219.
Grounding is by means of the ground terminal provided for the purpose inside the
connection housing. The stripped and twisted lengths of cable shield to the ground
terminal must be as short as possible.
7.1.3
Terminal assignment
Transmitter: supply voltage, input/outputs
The terminal assignment of the inputs and outputs depends on the individual order
version of the device. The device-specific terminal assignment is documented on an
adhesive label in the terminal cover.
Supply voltage
Input/output
1
Input/output
2
Input/output
3
Input/output
4
1 (+)
2 (–)
26 (B)
27 (A)
24 (+)
25 (–)
22 (+)
23 (–)
20 (+)
21 (–)
Device-specific terminal assignment: adhesive label in terminal cover.
Transmitter and sensor connection housing: connecting cable
The sensor and transmitter, which are mounted in separate locations, are interconnected
by a connecting cable. The cable is connected via the sensor connection housing and the
transmitter housing.
Terminal assignment and connection of the connecting cable: