Diagnostics and troubleshooting
Proline Promag H 500 PROFIBUS DP
162
Hauser
LED
Color
Meaning
5
Service interface (CDI),
Ethernet Link/Activity
Off
Not connected or no connection established.
Yellow
Connected and connection established.
Flashing yellow
Service interface active.
Proline 500
Different LEDs in the transmitter provide information on the device status.
1
2
3
4
5
A0029629
1
Supply voltage
2
Device status
3
Not used
4
Communication
5
Service interface (CDI) active, Ethernet Link/Activity
LED
Color
Meaning
1
Supply voltage
Off
Supply voltage is off or too low.
Green
Supply voltage is ok.
2
Device status
Off
Firmware error
Green
Device status is ok.
Flashing green
Device is not configured.
Red
A diagnostic event with "Alarm" diagnostic behavior
has occurred.
Flashing red
A diagnostic event with "Warning" diagnostic
behavior has occurred.
Flashing red/green
The device restarts.
3
Not used
–
–
4
Communication
Off
Device does not receive any Profibus data.
White
Device receives Profibus data.
5
Service interface (CDI),
Ethernet Link/Activity
Off
Not connected or no connection established.
Yellow
Connected and connection established.
Flashing yellow
Service interface active.
12.2.2 Sensor connection housing
Proline 500 – digital
Various light emitting diodes (LED) on the ISEM electronics (Intelligent Sensor Electronic
Module) in the sensor connection housing provide information on the device status.
Содержание Proline Promag H 500
Страница 240: ...www addresses endress com...