Proline Promag H 500 PROFIBUS DP
Commissioning
Hauser
111
10.6.4 Configuring the analog inputs
The
Analog inputs
submenu guides the user systematically to the individual
Analog input
1 to n
submenu. From here you get to the parameters of the individual analog input.
Navigation
"Setup" menu → Analog inputs
‣
Analog inputs
‣
Analog input 1 to n
Channel
PV filter time
Fail safe type
Fail-safe value
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Channel
–
Select the process variable.
• Volume flow
• Mass flow
• Corrected volume
flow
• Flow velocity
• Conductivity
• Corrected
conductivity
• Temperature
• Electronic
temperature
• Current input 1
• Current input 2
• Current input 3
Volume flow
PV filter time
–
Specify the time to suppress
signal peaks. During the
specified time the analog input
does not respond to an erratic
increase in the process
variable.
Positive floating-
point number
0
Fail safe type
–
Select the failure mode.
• Fail-safe value
• Fallback value
• Off
Off
Fail-safe value
In
Fail safe type
parameter,
the
Fail-safe value
option is
selected.
Specify the values to be output
when an error occurs.
Signed floating-point
number
0
*
Visibility depends on order options or device settings
10.6.5 Displaying the I/O configuration
The
I/O configuration
submenu guides the user systematically through all the parameters
in which the configuration of the I/O modules is displayed.