Diagnostics and troubleshooting
Proline Promag H 500 PROFIBUS DP
190
Hauser
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
530 Electrode cleaning is running
Turn off ECC
• Conductivity
• Corrected conductivity
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Good
Quality substatus
Function check
Coding (hex)
0xBC to 0xBF
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
531 Empty pipe adjustment faulty
Execute EPD adjustment
• Conductivity
• Corrected conductivity
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Corrected volume flow
• Volume flow
Measured variable status [from the factory]
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
537 Configuration
1. Check IP addresses in network
2. Change IP address
–
Measured variable status
Quality
Good
Quality substatus
Function check
Coding (hex)
0xBC to 0xBF
Status signal
F
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
594 Relay output simulation
Deactivate simulation switch output
–
Measured variable status
Quality
Good
Quality substatus
Function check
Coding (hex)
0xBC to 0xBF
Status signal
C
Diagnostic behavior
Warning