Proline Promag H 500 PROFIBUS DP
Commissioning
Hauser
137
Navigation
"Setup" menu → Advanced setup → Electrode cleaning circuit
‣
Electrode cleaning circuit
Electrode cleaning circuit
ECC duration
ECC recovery time
ECC cleaning cycle
ECC Polarity
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry / User
interface
Factory setting
Electrode cleaning circuit
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Enable the cyclic electrode
cleaning circuit.
• Off
• On
Off
ECC duration
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Enter the duration of electrode
cleaning in seconds.
0.01 to 30 s
2 s
ECC recovery time
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Define recovery time after
electrode cleaning. During this
time the current output values
will be held at last valid value.
1 to 600 s
60 s
ECC cleaning cycle
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Enter the pause duration
between electrode cleaning
cycles.
0.5 to 168 h
0.5 h
ECC Polarity
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Select the polarity of the
electrode cleaning circuit.
• Positive
• Negative
Depends on the
electrode material:
• Platinum:
Negative
option
• Tantalum, Alloy
C22, stainless
steel:
Positive
option
10.7.5 WLAN configuration
The
WLAN Settings
submenu guides the user systematically through all the parameters
that have to be set for the WLAN configuration.
Navigation
"Setup" menu → Advanced setup → WLAN Settings
‣
WLAN settings
WLAN IP address