Electrical connection
Proline Promag H 500 PROFIBUS DP
42
Hauser
7.1.6
Preparing the connecting cable: Proline 500 – digital
When terminating the connecting cable, pay attention to the following points:
‣
For cables with fine-wire cores (stranded cables):
Fit the cores with ferrules.
Transmitter
Sensor
A
70 (2.76)
50 (1.97)
10 (0.4)
8 (0.3)
B
1
A0029546
A
B
70 (2.76)
8 (0.31)
40(1.57)
1
A0029442
Engineering unit mm (in)
A = Terminate the cable
B = Fit ferrules on cables with fine-wire cores (stranded cables)
1 = Red ferrules, 1.0 mm (0.04 in)
7.1.7
Preparing the connecting cable: Proline 500
When terminating the connecting cable, pay attention to the following points:
1. In the case of the electrode cable:
Make sure that the ferrules do not touch the core shields on the sensor side.
Minimum distance = 1 mm (exception: green “GND” cable)
2. In the case of the coil current cable:
Insulate one core of the three-core cable at the level of the core reinforcement. You
only require two cores for the connection.
3. For cables with fine-wire cores (stranded cables):
Fit the cores with ferrules.
Содержание Proline Promag H 500
Страница 240: ...www addresses endress com...