Operation options
Proline Promag H 500 PROFIBUS DP
66
Hauser
8.2.2
Operating philosophy
The individual parts of the operating menu are assigned to certain user roles (operator,
maintenance etc.). Each user role contains typical tasks within the device lifecycle.
Menu/parameter
User role and tasks
Content/meaning
Language
task-oriented
Role "Operator", "Maintenance"
Tasks during operation:
• Configuring the operational
display
• Reading measured values
• Defining the operating language
• Defining the Web server operating language
• Resetting and controlling totalizers
Operation
• Configuring the operational display (e.g. display format, display contrast)
• Resetting and controlling totalizers
Setup
"Maintenance" role
Commissioning:
• Configuration of the
measurement
• Configuration of the inputs and
outputs
• Configuration of the
communication interface
Wizards for fast commissioning:
• Set the system units
• Display I/O/configuration
• Configure the inputs
• Configure the outputs
• Configuring the operational display
• Define the output conditioning
• Set the low flow cut off
• Configure empty pipe detection
Advanced setup
• For more customized configuration of the measurement (adaptation to
special measuring conditions)
• Configuration of totalizers
• Configuration of electrode cleaning (optional)
• Configure the WLAN settings
• Administration (define access code, reset measuring device)
Diagnostics
"Maintenance" role
Fault elimination:
• Diagnostics and elimination of
process and device errors
• Measured value simulation
Contains all parameters for error detection and analyzing process and device
errors:
• Diagnostic list
Contains up to 5 currently pending diagnostic messages.
• Event logbook
Contains event messages that have occurred.
• Device information
Contains information for identifying the device.
• Measured values
Contains all current measured values.
• Analog inputs
Is used to display the analog input.
•
Data logging
submenu with "Extended HistoROM" order option
Storage and visualization of measured values
• Heartbeat
The functionality of the device is checked on demand and the verification
results are documented.
• Simulation
Is used to simulate measured values or output values.