Proline Promag H 500 PROFIBUS DP
Electrical connection
Hauser
41
1
2
3
6
7
5
5
5
4
5
5
4
A0036639
1
Controller (e.g. PLC)
2
Cable shield
3
T-box
4
Measuring device
5
Local grounding
6
Bus terminator
7
Potential matching line
7.1.5
Preparing the measuring device
Carry out the steps in the following order:
1. Mount the sensor and transmitter.
2. Connection housing, sensor: Connect connecting cable.
3. Transmitter: Connect connecting cable.
4. Transmitter: Connect signal cable and cable for supply voltage.
NOTICE
Insufficient sealing of the housing!
Operational reliability of the measuring device could be compromised.
‣
Use suitable cable glands corresponding to the degree of protection.
1. Remove dummy plug if present.
2. If the measuring device is supplied without cable glands:
Provide suitable cable gland for corresponding connecting cable.
3. If the measuring device is supplied with cable glands: