System integration
Proline Promag H 500 PROFIBUS DP
104
Hauser
parameter via address shifting, the master can read out the current volume flow measured
value in slot 0 and index 230.
The data type (integer/float) and data access (read/write) depend on the parameter
entered in the configuration area. If the parameter entered supports read and write access,
the parameter can also be read- and write-accessed via the data area.