System integration
Proline Promag H 500 PROFIBUS DP
98
Hauser
Selection: totalizer value
Input variable
Volume flow
Mass flow
Corrected volume flow
Factory setting
Function block
Factory setting: TOTAL
Totalizer 1, 2 and 3
Volume flow
Data structure
Input data of TOTAL
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Measured value: floating point number (IEEE 754)
Status
SETTOT_TOTAL module
The module combination consists of the SETTOT and TOTAL functions:
• SETTOT: Control the totalizers via the PROFIBUS master.
• TOTAL: Transmit totalizer value, along with the status, to the PROFIBUS master.
Three Totalizer blocks are available (slot 5 to 7).
Selection: control totalizer
Value SETTOT
Control totalizer
0
Totalize
1
Resetting
2
Adopt totalizer initial setting
Factory setting
Function block
Factory setting: Value SETTOT (meaning)
Totalizer 1, 2 and 3
0 (totalizing)
Data structure
Output data of SETTOT
Byte 1
Control variable 1
Input data of TOTAL
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Measured value: floating point number (IEEE 754)
Status