Diagnostics and troubleshooting
Proline Promag H 500 PROFIBUS DP
194
Hauser
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
961 Electrode potential out of specification
1. Check process conditions
2. Check ambient conditions
• Empty pipe detection
• Low flow cut off
• Mass flow
• Volume flow
Measured variable status [from the factory]
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
962 Pipe empty
1. Perform full pipe adjustment
2. Perform empty pipe adjustment
3. Turn off empty pipe detection
• Conductivity
• Corrected conductivity
• Flow velocity
• Low flow cut off
• Mass flow
• Corrected volume flow
• Volume flow
Measured variable status [from the factory]
Quality
Bad
Quality substatus
Process related
Coding (hex)
0x28 to 0x2B
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
12.8 Pending diagnostic events
The
Diagnostics
menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via "FieldCare" operating tool → 168
• Via "DeviceCare" operating tool → 168
Other pending diagnostic events can be displayed in the
Diagnostic list
submenu
Navigation
"Diagnostics" menu
Diagnostics
Actual diagnostics
Previous diagnostics
Operating time from restart
Operating time