Electrical connection
Proline Promag H 500 PROFIBUS DP
46
Hauser
Connecting the sensor connection housing via terminals
For the device version with the order code for "Sensor connection housing":
Option
B
"Stainless, hygienic"
2.
1.
3.
6.
5.
10 (0.4)
8 mm
22 mm
7.
24 mm
A0029613
1. Release the securing screw of the housing cover.
2. Open the housing cover.
3. Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
4. Strip the cable and cable ends. In the case of stranded cables, fit ferrules.
5. Connect the protective ground.
6. Connect the cable in accordance with the connecting cable terminal assignment.
7. Firmly tighten the cable glands.
This concludes the process for connecting the connecting cable.
8. Close the housing cover.
9. Tighten the securing screw of the housing cover.