Proline Promag H 500 PROFIBUS DP
Diagnostics and troubleshooting
Hauser
177
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
275 I/O module 1 to n defective
Change I/O module
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
276 I/O module 1 to n faulty
1. Restart device
2. Change I/O module
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
283 Memory content
1. Reset device
2. Contact service
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm