Commissioning
Proline Promag H 500 PROFIBUS DP
140
Hauser
Options
Description
Compare
The device configuration saved in the device memory is compared with the current
device configuration of the HistoROM backup.
Clear backup data
The backup copy of the device configuration is deleted from the memory of the
device.
HistoROM backup
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
10.7.7 Using parameters for device administration
The
Administration
submenu systematically guides the user through all the parameters
that can be used for device administration purposes.
Navigation
"Setup" menu → Advanced setup → Administration
‣
Administration
‣
Define access code
‣
Reset access code
Device reset
Using the parameter to define the access code
Navigation
"Setup" menu → Advanced setup → Administration → Define access code
‣
Define access code
Define access code
Confirm access code
Parameter overview with brief description
Parameter
Description
User entry
Define access code
Restrict write-access to parameters to protect the configuration
of the device against unintentional changes.
Max. 16-digit character string comprising
numbers, letters and special characters
Confirm access code
Confirm the entered access code.
Max. 16-digit character string comprising
numbers, letters and special characters