Diagnostics and troubleshooting
Proline Promag H 500 PROFIBUS DP
186
Hauser
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
444 Current input 1 to n
1. Check process
2. Check current input settings
• Measured values 1
• Measured values 2
• Measured values 3
Measured variable status [from the factory]
Quality
Good
Quality substatus
Function check
Coding (hex)
0xBC to 0xBF
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
453 Flow override
Deactivate flow override
• Conductivity
• Corrected conductivity
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Good
Quality substatus
Function check
Coding (hex)
0xBC to 0xBF
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
463 Analog input 1 to n selection invalid
1. Check module/channel configuration
2. Check I/O module configuration
• Measured values 1
• Measured values 2
• Measured values 3
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
482 FB not Auto/Cas
Set Block in AUTO mode
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
F
Diagnostic behavior
Alarm