Diagnostics and troubleshooting
Proline Promag H 500 PROFIBUS DP
196
Hauser
F
I1091 Config. change
I1157 Mem.err. ev.list
F311 Electr. failure
/ ../Eventlist
0d01h19m10s
A0014008-EN
38
Taking the example of the local display
• A maximum of 20 event messages can be displayed in chronological order.
• If the
Extended HistoROM
application package (order option) is enabled in the device,
the event list can contain up to 100 entries .
The event history includes entries for:
• Information events → 196
In addition to the operation time of its occurrence, each event is also assigned a symbol
that indicates whether the event has occurred or is ended:
• Diagnostic event
– : Occurrence of the event
– : End of the event
• Information event
: Occurrence of the event
To call up the measures to rectify a diagnostic event:
• Via "FieldCare" operating tool → 168
• Via "DeviceCare" operating tool → 168
For filtering the displayed event messages → 196
12.10.2 Filtering the event logbook
Using the
Filter options
parameter you can define which category of event message is
displayed in the
Events list
submenu.
Navigation path
Diagnostics → Event logbook → Filter options
Filter categories
• All
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
• Information (I)
12.10.3 Overview of information events
Unlike a diagnostic event, an information event is displayed in the event logbook only and
not in the diagnostic list.
Info number
Info name
I1000
--------(Device ok)
I1079
Sensor changed
I1089
Power on
I1090
Configuration reset
I1091
Configuration changed