Diagnostics and troubleshooting
Proline Promag H 500 PROFIBUS DP
172
Hauser
12.7 Overview of diagnostic information
The amount of diagnostic information and the number of measured variables affected
increase if the measuring device has one or more application packages.
In the case of some items of diagnostic information, the diagnostic behavior can be
changed. Change the diagnostic information → 169
12.7.1 Diagnostic of sensor
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
043 Sensor short circuit
1. Check sensor cable and sensor
2. Execute Heartbeat Verification
3. Replace sensor cable or sensor
• Conductivity
• Corrected conductivity
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status [from the factory]
Quality
Uncertain
Quality substatus
Maintenance demanded
Coding (hex)
0x68 to 0x6B
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
082 Data storage
1. Check module connections
2. Contact service
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
083 Memory content
1. Restart device
2. Restore HistoROM S-DAT backup
('Device reset' parameter)
3. Replace HistoROM S-DAT
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm