| Appendix A: Troubleshooting
Microplate Carrier Movement
Problem
Possible Cause
Possible Solution
Microplate not properly
seated or strips not
level.
Reseat microplate in carrier or strips
in holder.
Make sure the carrier is clean (see
Cleaning the Plate Carrier System
in
Ch. 4
).
Try a different microplate or strip
holder. If the problem is unresolved,
the carrier may have to be realigned.
Contact Bio-Tek TAC.
Horizontal Aspirate
Position is programmed
too wide for a
movement.
Change the Horizontal Aspir Pos
value in the program (see
Define
Wash ASPIRATION Program
Options
in
Ch. 3
).
Aspiration
tubes not
entering
well
correctly.
Aspirate tubes bent.
Contact Bio-Tek TAC.
Washer Manifold Movement
Problem
Possible Cause
Possible Solution
Manifold movement is
blocked.
Check orientation of microplate; A1
should be in the left rear corner of
the plate carrier as you face the front
of the instrument.
Check for and remove any
obstructions.
Ensure that the manifold is installed
properly (see
Setting Up the
ELx405: Installing the Microplate
Carrier
in
Ch. 2
).
Manifold
position
error.
Incorrect manifold
selected.
HT models with the accessory 96-
tube manifold
(including Select
models upgraded to HT models):
Ensure that the washer is configured
for the installed manifold (96-tube or
192-tube). See
Manifold Selection
in
Ch. 3
.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...