Liquid Tests
|
165
Dispense Precision Test Worksheet
for Single/96 and Dual/96 Manifolds
without the External Valve Module
Washer Model:
ELx405™
Select CW
Magna
HT2 (with Dual/96 manifold)
Select
Serial Number:
Calculations
Standard Deviation:
(calculate using spreadsheet program)
Mean OD:
(sum of all wells
÷
number of wells)
% Coefficient of Variation:
((Standard Deviation
÷
Mean OD) x 100)
% CV <= 4.0?
Pass Fail
Date:
Test Performed By: