60
| Chapter 3: Operation
Wash DISPENSE (DISP) Defaults and Ranges, cont’d
Parameter
Units
Default
Range
Models
-30 to 30
ELx405 or Magna
-45 to 45
with 96-well
plates
Select, Select CW,
HT/96-tube
Horizontal
Dispense Position
quarter
steps
0
-25 to 25
with 384-
well plates
Select, Select CW,
HT/96-tube
HT/192-tube
-20 to 20
with 96-well
plates
Select, Select CW
HT/96-tube
Horizontal Y
Dispense Position
(Select, Select CW,
HT only)
quarter
steps
0
-10 to 20
with 384-
well plates
Select, Select CW
HT/96-tube
HT/192-tube
Bottom Wash
First?
No
No/Yes
Bottom
Dispense Volume
µ
l/well
300
50-3000
Bottom Flow Rate
5
1-9
Bottom
Dispense Height
60
12-180
All
Notes:
•
The shaded options above are
not
available for Dispense-only programs.
•
Horizontal Dispense Position
and
Horizontal Y Dispense Position
: For
higher throughput of the HT with 384-well plates, use the 192-tube manifold.
•
Bottom Dispense Height
: Although the range is 12-180, the recommended
highest value is
175
.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...