Troubleshooting Charts
|
181
Fluid Leakage
Problem
Possible Cause
Possible Solution
Defective seals.
Replace individual seals at the front
of the manifold, or o-rings at the rear
of the manifold (see
Cleaning the
Manifold
in
Ch. 4
).
Replace o-rings on washer manifold
inlet fittings.
Contact Bio-Tek TAC.
Fluid
leaking
from
manifold.
Aspiration tubes only:
vacuum too low.
Check waste connector tubes; make
sure they are properly connected to
the manifold.
If you are using an in-line vacuum
filter, check the filter for clogging,
and replace if necessary.
Check seal of waste bottle covers.
Check for air leaks in the waste
tubing and bottles.
Use a slower Aspiration Rate in the
program (see
Define Wash
ASPIRATION Program Options
in
Ch. 3
).
Defective tubing
connector or inlet
tubing.
Contact Bio-Tek TAC.
Fluid
leaking
from
underneath
the washer.
Leaking valve.
Contact Bio-Tek TAC.
Defective connector.
Replace connector.
Worn tubing.
Replace tubing, or cut back tubing
one inch (to remove worn section).
Fluid
leaking
from
external
tubing
connector.
Worn seal (inlet or
vacuum fitting).
Replace filter or seal (see
Connecting the Tubes and Bottles
in
Ch. 2
).
ELx405 Operator's Guide
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...