Troubleshooting Charts
|
183
Microplate Scratches
Problem
Possible Cause
Possible Solution
Microplate dispense or
aspiration height
adjustment too low.
Change the Dispense or Aspiration
Height in the program (see
Define
Wash DISPENSE or ASPIRATION
Program Options
in
Ch. 3
).
Scratches
on
microplate
bottom.
Microplate not properly
seated or strips not
level.
Reseat microplate in carrier or strips
in holder.
Make sure the carrier is clean (see
Cleaning the Plate Carrier System
in
Ch. 4
).
Try a different microplate or strip
holder. If the problem is unresolved,
the carrier may have to be realigned.
Contact Bio-Tek TAC.
ELx405 Operator's Guide
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...