Prepare for Storage or Shipment
|
135
6.
If the washer is equipped with the
external valve module
, connect the
supply bottles this way:
Valve A: Disinfectant bottle
Valve B: Rinse solution bottle
Valve C: Empty bottle
7.
Turn on the washer.
8.
To run the
LONG_SHUTDOWN
program, follow this sequence:
While this program is running, you will need to periodically check the
display panel and follow the instructions.
Washers not equipped with the
external valve module:
MAINT
SELECT MAINT PROGRAM:
LONG_SHUTDOWN
CONNECT DISINFECTANT AND PRESS
<START>
KEY
Washers equipped with the
external valve module:
MAINT
SELECT MAINT PROGRAM:
LONG_SHUTDOWN
DISINFECTANT VALVE?
A
RINSE VALVE?
B
AIR VALVE?
C
START MAINTENANCE?
YES
Storing the Washer
After performing the
Long_Shutdown
procedure:
1.
Turn off the washer and disconnect the power cord.
2.
Store the washer on a flat surface that is relatively free of vibration, in a
dust-free and particle-free environment.
3.
Protect the washer from temperature extremes that can cause condensation
within the unit and from corrosive fumes and vapors.
4.
Store the washer under the following environmental conditions:
Temperature:
–20° to 50°C (-4° to 122°F)
Relative humidity:
10% to 85% (non-condensing)
ELx405 Operator's Guide
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...