Appendix B: Error Codes
Bio-Stack Errors,
Cont’d
Code
Cause
B-8305
INVALID_RESPONSE_ERR
The ELx405 has received an invalid message
response from the Bio-Stack. Occurs when
neither a NAK nor an ACK is received from
the Bio-Stack. This can happen during any
message sent to the Bio-Stack, including
those associated with retrieving information,
as well as those used for moving motors.
Probable Causes:
•
The ELx405 was turned on before the
Bio-Stack finished its startup routine.
•
The ELx405 was turned on first, then
the Bio-Stack was turned on.
•
The Bio-Stack has an error or is in a
bad state and needs to be turned off
and then turned back on.
•
The cable connection is lost, not
allowing communications to transfer
properly.
B-8306
TIMEOUT_ERR
The ELx405 failed to communicate with the
Bio-Stack because the serial communications
timed out. This can happen during any
message sent to the Bio-Stack, including
those associated with retrieving information,
as well as those used for moving motors.
Probable Causes:
•
The ELx405 was turned on before the
Bio-Stack finished its startup routine.
•
The ELx405 was turned on first, then
the Bio-Stack was turned on.
•
The Bio-Stack has an error or is in a
bad state and needs to be turned off
and then turned back on.
•
The cable connection is lost, not
allowing communications to transfer
properly.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...