Rinse/Soak the Fluid Path
|
113
3
Connect a bottle containing four liters of deionized water and run
the
P_DAY_RINSE
Prime program three times to flush the system.
Using Ethanol:
At the end of the day:
1
Prepare a four-liter solution of 10% ethanol and deionized water.
(
Do not
use isopropyl alcohol.) Connect the bottle to the washer
(Valve “A” if the valve module is connected).
2
Run the
DECONTAMINATION
Maintenance program.
3
Connect a bottle containing four liters of deionized water and flush
the system by running either the
P_DAY_RINSE
Prime program
(three times), or the
OVERNIGHT_LOOP
Maintenance program.
Using Deionized Water:
If the washer will be idle between plates for longer than 45 minutes, flush
the proteins from the washer:
1
Fill a supply bottle with deionized water. Connect the bottle to the
washer (Valve “A” if the valve module is connected).
2
Run the
P_DAY_RINSE
Prime program.
3
Enable
AUTOPRIME
for 60-minute intervals.
At the end of the day:
1
Fill a supply bottle with deionized water. Connect the bottle to the
washer (Valve “A” if the valve module is connected).
2
Run the
P_DAY_RINSE
Prime program three times.
3
Run the
OVERNIGHT_LOOP
Maintenance program.
Weekly/Monthly
1
Flush the system with 0.1-0.5 N NaOH, followed by neutralization
with an equivalent normality (0.1-0.5 N) of HCl.
2
Rinse well with deionized water to remove the HCl.
3
Run the
OVERNIGHT_LOOP
Maintenance program with deionized
water, or run the
P_DAY_RINSE
Prime program three times with
deionized water if you plan to use the washer immediately.
ELx405 Operator's Guide
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...