| Chapter 2: Installation
Before Repackaging the Instrument
Warning!
If the washer has been exposed to potentially
hazardous material,
decontaminate it
to minimize the risk to all
who come in contact with the washer during shipping, handling
and servicing.
Decontamination prior to shipping is required by the U.S.
Department of Transportation regulations.
Important!
Perform the following steps in the order presented
before repackaging the washer for shipment.
Important!
Failure to fasten the manifold in place with the
shipping bracket before shipment could result in irreparable
manifold damage.
1.
Decontaminate
the washer and its accessories as necessary. Refer to the
Decontamination
procedure in
Chapter 4, Preventive Maintenance
for
complete instructions.
2.
HT2/HTV2 models:
If the 192-tube manifold is not installed, install it
now. Store the 96-tube manifold in its special case.
3.
Run the
PARK
utility.
PARK
prepares the manifold for installing the
manifold shipping bracket.
UTIL
SETUP
MORE
PARK
4.
Turn the washer off and disconnect the power cord and all tubing.
5.
Remove the mist shield and carrier, and set aside for packing. Screw the
mist shield mounting screws back into the base.
6.
Reinstall the
manifold shipping bracket
(see
Figure 5
on page 18 and
Figure 8
on page 21).
7.
Obtain a
Return Materials Authorization
(RMA) number from
Bio-Tek’s Technical Assistance Center through Bio-Tek’s website, fax, or
e-mail address listed in
Chapter 1
.
When obtaining the RMA, explain whether the washer requires calibration,
cleaning, periodic maintenance, warranty work, and/or repair. Make a note
of any error messages displayed and their frequency.
Provide Bio-Tek with the name and contact information of a person who
may be contacted if questions arise.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...