Fatal Errors
|
201
Fatal Errors,
Cont’d
Code
Cause
A302
NOT_AVAIL_ERR
Magna models only:
Carrier x-axis motor not
available – occurs when trying to get access to
a system resource, such as the carrier x-axis
or a timer. If the requested device is the x-axis
motor, usually that motor is still running, or
the software “thinks” the motor is still running,
even if it isn’t.
A303
NOT_AVAIL_ERR
Dual manifold models only:
Carrier y-axis
motor not available – occurs when trying to get
access to a system resource, such as the
carrier y-axis motor or a timer. If the
requested device is the y-axis motor, usually
that motor is still running, or the software
“thinks” the motor is still running, even if it
isn’t.
A304
NOT_AVAIL_ERR
Software timer not available.
A305
NOT_AVAIL_ERR
Display not available.
A306
NOT_AVAIL_ERR
Quick flash memory not available.
A307
NOT_AVAIL_ERR
Memory manager (memory allocation heap)
not available.
A400
CHECKSUM_ERR
Failed code checksum test on powerup –
returned by EPROM software when
downloading code unsuccessfully, or when it is
trying to initialize after a successful download.
A500
POWER_ERR
Power dropped below safe level – occurs at
powerup and when trying to perform a write to
flash memory, if there is insufficient power to
perform the desired operation.
A600
QFLASH_TIMEOUT_ERR
Quick flash memory configuration timed out.
Occurs in the interrupt handler routine, which
is triggered every 1 msec, if there is a quick
flash memory error.
ELx405 Operator's Guide
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...