| Appendix A: Troubleshooting
Fluid Delivery
Problem
Possible Cause
Possible Solution
Clogged fluid inlet filter.
Clean the fluid inlet filter. See
Cleaning the Fluid Inlet Filter,
in
Ch. 4
; note orientation of the filter in
Figure 28
.
Supply tube inside the
supply bottle is kinked
or disconnected.
Straighten or connect supply tube.
Make sure the end is cut; see
General
Usage Guidelines
,
Ch. 3
.
Inlet tube not
connected.
Check all tubing (see
Connecting the
Tubes and Bottles
in
Ch. 2
).
Clogged dispense tubes
on the washer manifold.
Remove and clean the manifold (see
Cleaning the Manifold
in
Ch. 4
).
No wash or rinse fluid.
Fill bottles with appropriate fluid.
Ensure bottles are clean and do not
contain particles or organic material.
System not primed.
Large air pockets in
tubing.
Run a Prime program using the
following volumes;
do not
use the
AutoPrime program for the initial
prime:
All models except Select CW
: 200
ml Prime Volume when all the lines are
empty or when changing fluids.
Select CW models
: 200 ml Prime
Volume and 150 ml Low Flow Prime
Volume when all the lines are empty;
300 ml Prime Volume and 200 ml Low
Flow Prime Volume when changing
fluids.
Unable to
dispense
fluid:
ELx405
models
without
the
external
valve
module.
Insufficient suction
force, clogged tubing,
or faulty valve.
Perform Periodic Maintenance as
described in
Ch. 4
. If problem
persists, contact Bio-Tek TAC.
Unable to
dispense
fluid:
ELx405
models
with
the external
valve
module.
System not primed.
Large air pockets in
tubing.
Run a Prime program using the
following volumes:
All models except Select CW
: 400
ml Prime Volume when all the lines are
empty or when changing fluids.
Select CW models
: 400 ml Prime
Volume and 300 ml Low Flow Prime
Volume when all the lines are empty;
600 ml Prime Volume and 400 ml Low
Flow Prime Volume when changing
fluids.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...