Wash
|
57
Define Wash DISPENSE (DISP) Program Options, Cont’d
Option & Description
Models
Action/Comments
ELx405,
Magna
The range is –30 to 30.
Select,
Select CW,
HT/96-tube
The range is –45 to 45 with 96-well
plates.
Select,
Select CW,
HT/96-tube,
HT/192-tube
The range is –25 to 25 with 384-well
plates.
Note:
For higher throughput of the HT
with 384-well plates, use the 192-tube
manifold.
Horizontal Disp Pos
Horizontal Dispense
Position represents the
left and right position of
the dispense tubes when
the carrier is beneath the
manifold.
The washer display shows the corresponding
measurement in millimeters for each option.
A
negative
offset positions the dispense tubes toward
the left side of the well. A
positive
offset positions the
dispense tubes toward the right side of the well. The
default setting of 0 indicates no offset (0.000 mm).
Note:
A more negative offset combined with a low flow
rate, are the recommended settings for “gentle washing,”
such as required by cell-based assays.
To visually determine the best offset parameters for a
microplate or assay, use the
Adjust Utility
(see page
92).
Select,
Select CW,
HT/96-tube
The range is –20 to 20 with 96-well
plates.
Select,
Select CW,
HT/96-tube,
HT/192-tube
The range is –10 to 20 with 384-well
plates.
Note:
For higher throughput of the HT
with 384-well plates, use the 192-tube
manifold.
Horiz Y Disp Pos
Horizontal Y Dispense
Position represents the
carrier Y axis position
(front/back movement)
which aligns the
microplate wells with the
manifold tubes during a
dispense.
The washer display shows the corresponding
measurement in millimeters for each option.
A
negative
offset positions the dispense tubes toward
the front side of the well.
A
positive
offset positions the dispense tubes away from
the front side of the well.
A setting of 0 indicates no offset (0.000 mm).
To visually determine the best offset parameters for a
microplate assay, use the
Adjust Utility
(see page 92).
ELx405 Operator's Guide
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...