Verifying Performance
|
35
Verifying Performance
Before using the ELx405
Washer for the first time, verify that it is operating properly
by turning the washer on. The washer will automatically perform a System Self Test to
check the manifold, and manifold-to-carrier positioning. It also performs a magnetic
carrier load test on the Magna models.
You can also choose to run this test from the washer's front panel. To run the washer’s
System Self Test, follow the washer menu path shown below.
UTIL
TESTS
SLFCHK
If the test passes
, the Main Menu will be displayed.
If the test fails
, the washer will beep repeatedly and the display will show an error
code. If this happens, note the error code and then press the
Stop
key on the keypad to
stop the beeping. Refer to
Appendix B
,
Error Codes
, to determine its cause. If the
problem is something you can fix, turn off the washer, fix the problem, and then turn
the washer back on. If the cause is not something you can fix, contact Bio-Tek’s
Technical Assistance Center.
Chapter 5
,
Performance Verification/Qualification Tests
provides recommended
Installation and Operational Qualification procedures to be performed after the
instrument is installed and set up as described in this chapter, and
before
the
instrument is used in a laboratory environment.
The successful completion of the Installation Qualification confirms that the washer
and its components have been supplied as ordered and ensures that they are
assembled and configured properly for your lab environment.
The successful completion of the Operational Qualification confirms that the washer is
operating according to specification.
Note:
An
instrument qualification package
(PN 7100567) for the ELx405 is
available for purchase from Bio-Tek. The package contains thorough procedures for
performing Installation Qualification, Operational Qualification and Performance
Qualification (IQ-OQ-PQ) and preventive maintenance (PM). Extensive Checklists and
Logbooks are included for recording results. Contact your local dealer for more
information.
Important!
Before operating this washer, review the
General Usage
Guidelines
in
Chapter 3
,
Operation
. These guidelines include
necessary steps to perform before running any program, information
on performing periodic maintenance, and points to consider when
creating or editing washer programs.
ELx405 Operator's Guide
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...