194
|
Appendix B: Error Codes
General Errors,
Cont’d
Code
Cause
1900
MALLOC_ERR
Memory allocation failure - occurs when trying to store
a large block of configuration data in flash memory,
such as downloaded configuration data, or autocal
configuration data.
Probable Causes:
•
The memory is corrupt. Contact Bio-Tek TAC.
•
If the error persists, the processor PCB needs to
be replaced.
1A00
ASPIRATE_POS_ERR
Horizontal aspirate position is out of range – occurs
when verifying the validity of a program prior to
running the program, or prior to saving the program to
flash memory.
Probable Causes:
•
The horizontal aspirate position requested is out
of range. See
Wash ASPIRATE Defaults and
Ranges
in
Chapter 3
,
Operation
for a list of
acceptable ranges.
1B00
DISPENSE_POS_ERR
Horizontal dispense position is out of range – occurs
when verifying the validity of a program prior to
running the program, or prior to saving the program to
flash memory.
Probable Causes:
•
The horizontal dispense position requested is
out of range. See
Wash DISPENSE Defaults
and Ranges
in
Chapter 3
,
Operation
for a list
of acceptable ranges.
1E00
AUTOCAL_CHECKSUM_ERR
Failed autocal checksum during the powerup sequence;
autocal data is missing.
Probable Causes:
•
Basecode is loaded, but autocalibration has not
been performed.
•
Memory corruptions.
•
Autocalibration has not been performed.
•
Main PCB was replaced.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...