162
| Chapter 5: Performance Verification/Qualification Tests
c. Calculate
the
%CV:
(Standard Deviation / Mean OD) * 100.
The %CV should be
<=
4.0
.
8.
If the %CV is
greater than
4.0, one or more dispense tubes may need to be
cleaned. Run
AUTOCLEAN
(Ultrasonic Advantage™ models only), and/or
remove the manifold and use the stylus to clean the dispense tube(s) giving
lower-than-average absorbance readings (all models). When finished, re-
prime the washer and retry the test.
9.
When finished, prime with deionized water to flush out the dye solution.
Annual Procedure for All Valves
Perform this test during the
annual OQ
only.
The
routine
test for Valve A must pass before the
annual
test for
Valves A-D can be performed.
1.
Empty the waste bottle now, and then as needed throughout this
procedure.
2.
Fill each of the supply bottles connected to
Valves A, B, C,
and
D
with
three liters (3000 ml) of deionized water.
Repeat the following steps for each valve:
3.
Run the Maintenance program
DAY_RINSE
two or three times to prime
the fluid lines, manifold, and the valve being tested.
4.
Edit the Dispense program
192accuracy_qc
to use the Valve currently
being tested (
DEFINE
EDIT
DISP
192accuracy_qc
).
5.
Place a clean, dry 384-well microplate on the balance and zero the balance.
6.
Place the microplate on the washer and run the Dispense program
192accuracy_qc
. This program dispenses 80
µ
l of water to each well of
the plate. It does not evacuate the wells.
7.
When the program is finished, carefully remove the plate and weigh it. This
is the
Total Dispense Weight
in grams.
The weight should be
30.72 grams, ± 10%
(between 27.65 g and
33.79 g).
If the weight falls outside this range, contact Bio-Tek.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...