Liquid Tests
|
169
Dispense Precision Test Worksheet
for the Dual/192 Manifold
with the External Valve Module
Washer Model:
HT
HTV2 (with Dual/192 manifold)
Serial Number:
Calculations for Valve A
Standard Deviation:
Mean OD (Sum of all wells
÷
Number of wells):
% CV (Standard Deviation
÷
Mean OD x 100):
% CV <= 4.0?
Pass Fail
Calculations for Valves A-D
(Annual OQ only, check here
if not performed)
Total Dispense Weight
30.72 g ±10%?
(27.65 g – 33.79 g)
Valve A
grams
Pass Fail
Valve B
grams
Pass Fail
Valve C
grams
Pass Fail
Valve D
grams
Pass Fail
Date:
Test Performed By: