Chapter 4: Preventive Maintenance
Tip:
Detergent such as Terg-A-Zyme
®
added to deionized
water in the supply bottle helps to break down the water’s
surface tension and enhances the cleaning process.
Terg-A-Zyme
®
also contains protease enzyme for assimilating
protinaceous residue such as bovine serum albumin (BSA).
Run one or more loops in the cleaning program for 30
minutes in each loop. Follow with a rinse loop using deionized
water to remove the detergent from the system, or with a
wash buffer to leave the instrument primed and ready for
use.
Prepare the waste and supply bottles
Empty the waste bottle. Ensure that there is a
sufficient volume
of
deionized water and detergent in the supply bottle(s): An
AUTOCLEAN
program initially primes the system with
200 ml
of fluid from the supply
bottle specified for Loop 1; each cleaning loop fills the reservoir with
approximately
93 ml
of fluid.
Priming with 200 ml is always done for
each successive
loop
in models without the valve module, and in models with
the valve module, if the valve selection changes between
loops.
Define the AUTOCLEAN program parameters:
From the washer’s main menu, follow the path below:
Washers not equipped with the
external valve module
:
UTIL
AUTCL
DEFINE
AUTOCLEAN NUM LOOPS: Enter
01
to
10
loops
LOOP <
loop number
> DURATION: Enter
01
to
60
minutes
OK TO SAVE PROGRAM?
YES
Washers equipped with the
external valve module
:
UTIL
AUTCL
DEFINE
AUTOCLEAN NUM LOOPS: Enter
01
to
10
loops
LOOP <
loop number
> DURATION: Enter
01
to
60
minutes
LOOP <
loop number
> VALVE: Select
A
,
B
,
C
, or
D
OK TO SAVE PROGRAM?
YES
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...