116
|
Chapter 4: Preventive Maintenance
4.
If detergent was used, wipe the components with a cloth moistened
with water. Use a clean, dry cloth to dry the components.
5.
To replace the carrier, line up the pin on the underside of the carrier
with the slot on the carrier transport rail. Snap the two carrier rail
guides onto the rail. The pin should sit in the slot.
Cleaning the Carrier Transport Arm (Select, Select CW, and HT
Models)
For troubleshooting purposes, it may be necessary to disassemble and clean the
spring-loaded carrier transport arm that allows the Select, Select CW, and HT
models to move forward and back (along the y-axis) when washing 384-well
plates.
If reagent has spilled and dried on the stainless steel shaft of the arm, it may
interfere with carrier movement. Cleaning the arm will resolve the problem.
Figure 22: Position of the Transport Arm on the Plate Carrier
The transport arm is noted with an arrow in the photo above. As shown in the
photo, the arm fits into the left side of the plate carrier.
Follow the instructions on the next two pages for removing, cleaning, and
reassembling the transport arm.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...