Chapter 4: Preventive Maintenance
Maintenance Program:
RINSE_AND_SOAK
Prime Program:
P_RINSE_AND_SOAK
Parameter Default
Selectable Range
Prime Volume
400 ml (All models except Select CW);
300 ml (Select CW)
1-999 ml
Low Flow Prime Volume
(Select CW models only)
200 ml
1-999 ml
Flow Rate
7
1-9
Soak After Prime?
Yes
Yes/No
Soak Duration
5 minutes
1 minute to 18 hours
Maintenance Program:
LONG_SHUTDOWN
and
DECONTAMINATION
Prime Program:
P_DECON1
Parameter Default
Selectable Range
Prime Volume
400 ml (All models except Select CW);
300 ml (Select CW)
1-999 ml
Low Flow Prime Volume
(Select CW models only)
200 ml
1-999 ml
Flow Rate
7
1-9
Soak After Prime?
Yes
Yes/No
Soak Duration
20 minutes
1 minute to 18 hours
Maintenance Program:
LONG_SHUTDOWN
and
DECONTAMINATION
Prime Program:
P_DECON2
Parameter Default
Selectable Range
Prime Volume
800 ml (All models except Select CW);
600 ml (Select CW)
1-999 ml
Low Flow Prime Volume
(Select CW models only)
200 ml
1-999 ml
Flow Rate
7
1-9
Soak After Prime?
Yes
Yes/No
Soak Duration
2 minutes
1 minute to 18 hours
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...