Liquid Tests
|
159
If the problem appears to be related to particular wells, clean the
associated aspiration tubes: run
AUTOCLEAN
(Ultrasonic Advantage™
models only), and/or remove the manifold and thoroughly clean the
tubes with the stylus (all models). See
Chapter 4
,
Preventive
Maintenance
for complete instructions. When finished, retry the test.
If the problem appears to be related to a particular region, edge, or
corner of the plate, review the alignment and flatness of the plate on the
carrier.
Please do not adjust
the carrier adjustment screws! Contact
your Bio-Tek dealer if you suspect an alignment problem.
For additional suggestions, refer to
Appendix A
,
Troubleshooting
.
If the test continues to fail, contact Bio-Tek Instruments.