Liquid Tests
|
167
Evacuation Efficiency Test Worksheet
for the Dual/192-Tube Manifold
Washer Model:
HT
HT2/HTV2 (with Dual/192 manifold)
Serial Number:
Standard Test
Total Residual Weight: g
Verification that wells are consistent in appearance:
Pass Fail
Mean Residual Weight (Total Residual Weight
÷
384): g
Mean Residual Weight <= 0.002 g?
Pass Fail
Supplemental Test (check here if not performed)
Mean OD for the plate (sum of all wells
÷
384):
Residual Factor (Mean OD
÷
Mean Residual Weight):
Calculate the Residual Weight for each well: well OD
÷
Residual Factor
Every Residual Weight per well <= 0.002 g?
Pass Fail
Date:
Test Performed By:
ELx405 Operator's Guide
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...