| Chapter 3: Operation
Adjust Utility
(ADJUST)
The
ADJUST
utility
allows the user to view the positions of the aspirate and
dispense tubes in relation to the microwells, and “single-step” the carrier or
manifold axes through their offset ranges to select the best offset. For example, use
the Adjust utility to determine the most effective setting for the Horizontal
Dispense Position in a Wash program.
1.
Place a microplate on the carrier.
2.
To view aspirate/dispense tube positions or adjust an offset, follow the
menu path below.
UTIL SETUP ADJUST
PLATE TYPE
ADJUSTMENT POSITION AXIS
3.
From the
PLATE TYPE
screen, choose
96
or
384
for the Select, Select CW,
and HT/96-tube manifold, and
96
or
96Bead
for the Magna. (
PLATE TYPE
does not appear in the standard ELx405 or HT with the 192-tube manifold
installed.)
4.
At the
ADJUSTMENT POSITION
screen, select a tube position:
ASPIR
(aspiration),
DISP
(dispense), or
MAGNET
(magnetic plate – Magna model
only). Only one tube position may be viewed at a time.
5.
At the
AXIS
selection screen, choose an axis. The top line of the display
indicates which axis is active, and the offset position of that axis. The axis
options are model-dependent, as follows:
MAN
Manifold (up/down movement)
All models
CARX
Carrier x-axis (left/right movement)
All models
CARY
Carrier y-axis (front/back movement)
Select, Select CW, HT
MAGNET
Magnet plate carrier (up/down
movement)
Magna
6.
Press the
◄
(reverse) key to single-step the offset in a negative direction,
and the
►
(forward) key to single-step the offset in a positive direction. See
the examples in
Figure 21
on the following page.
7.
When the desired offset position is found, record the position number for
later use when selecting an offset for a wash, dispense, or aspirate program.
8.
To quit the Adjust utility, press
Main Menu
. When the selection is made,
the software automatically repositions the carrier and manifold to their
default
positions.
Bio-Tek Instruments, Inc.
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...