Warnings & Precautions
|
105
Warnings & Precautions
Please read the following before performing any Maintenance procedures:
Warning! Internal Voltage.
Turn off and unplug the
instrument for all maintenance and repair operations.
Caution!
Chemical Compatibility with Washers.
Some chemicals may cause irreparable damage to washers.
The following chemicals have been deemed safe for use in
washers: buffer solutions (such as PBS), saline, surfactants,
deionized water, 70% ethyl, isopropyl, or methyl alcohol,
40% formaldehyde, and 20% sodium hydroxide.
Never use DMSO or other organic solvents. These chemicals
may cause severe damage to the instrument. See
Appendix
C
,
Chemical Compatibility
, for more information. Use of
wash buffers containing acetic acid is limited to washers
upgraded with Bio-Tek part number 68098 Teflon
®
valves.
Contact Bio-Tek for upgrade information and prior to using
other questionable chemicals.
Warning!
Wear protective gloves when handling
contaminated instruments. Gloved hands should be
considered contaminated at all times; keep gloved hands
away from eyes, mouth, nose, and ears.
Warning!
Mucous membranes are considered prime entry
routes for infectious agents. Wear eye protection and a
surgical mask when there is a possibility of aerosol
contamination. Intact skin is generally considered an effective
barrier against infectious organisms; however, small
abrasions and cuts may not always be visible. Wear
protective gloves when handling contaminated instruments.
Important!
Do not immerse the instrument, spray it with
liquid, or use a “wet” cloth on it. Do not allow water or other
cleaning solution to run into the interior of the instrument. If
this happens, contact Bio-Tek’s Technical Assistance Center.
ELx405 Operator's Guide
Содержание ELx405
Страница 2: ......
Страница 3: ...ELx405 Microplate Washer Operator s Guide January 2005 2005 7101000 Revision I Bio Tek Instruments Inc...
Страница 4: ...ii Preface Bio Tek Instruments Inc...
Страница 6: ...iv Preface Bio Tek Instruments Inc...
Страница 70: ...40 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 157: ...Clean Components 127 Vacuum port Fluid inlet filter Figure 28 Cleaning the Fluid Inlet Filter ELx405 Operator s Guide...
Страница 168: ...138 Chapter 4 Preventive Maintenance Bio Tek Instruments Inc...
Страница 200: ...170 Chapter 5 Performance Verification Qualification Tests Bio Tek Instruments Inc...
Страница 214: ...184 Appendix A Troubleshooting Bio Tek Instruments Inc...
Страница 250: ...220 Index Bio Tek Instruments Inc...